RecallRadar

FDA Device recall Z-0571-2025

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

On 2024-12-04 the FDA classified a Class II recall of Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software by Livanova Deutschland, citing heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, bu….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0571-2025
ClassificationClass II
Statusongoing
Initiated2024-10-18
Published2024-12-04
Last updated2026-09-13 11:57 UTC
CompanyLivanova Deutschland
DistributionUS

Product

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Reason

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Identifiers

code_info
UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 4…UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0571-2025%22