On 2012-12-26 the FDA classified a Class II recall of Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086 by Medtronic Neuromodulation, citing medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0567-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-12-03 |
| Published | 2012-12-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Medtronic Neuromodulation |
| Distribution | US |
Product
Medtronic, DBS Extension Kit for Deep Brain Stimulation, Catalog # 7482, 37085, 37086. Rx Only, Manufactured in: Medtronic, Inc. Villalba, Puerto Rico. Medtronic¿ DBSTM Therapy for Parkinson's Disease - Bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) using Medtronic DBS Therapy for Parkinson's Disease is indicated for adjunctive therapy in reducing some of the symptoms of advanced, levodopa-responsive Parkinson's disease that are not adequately controlled with medication. STN/GPi target system components: ¿ Kinetra¿ Model 7428 or Soletra¿ Model 7426 Neurostimulator ¿ Activa¿ PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead Medtronic DBS Therapy for Tremor - Unilateral thalamic stimulation of the ventral intermediate nucleus (VIM) using Medtronic DBS Therapy for Tremor is indicated for the suppression of tremor in the upper extremity. The system is intended for patients who are diagnosed with essential tremor or parkinsonian tremor not adequately controlled by medications and where the tremor constitutes a significant functional disability. VIM target system components: ¿ Soletra Model 7426 Neurostimulator ¿ Activa PC Model 37601, Activa RC Model 37612, Activa SC Model 37602, or Activa SC Model 37603 Neurostimulator ¿ Model 7482, Model 7482A, Model 7483, Model 37085, or Model 37086 Extension ¿ Model 3387 or Model 3389 Lead.
Reason
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
Identifiers
| code_info | NHU236350V to NHU236373V; NKN037257V to NKN037259V; NKN037372V to NKN037387V; NKN037476V to NKN037507V; NKN037692V…NHU236350V to NHU236373V; NKN037257V to NKN037259V; NKN037372V to NKN037387V; NKN037476V to NKN037507V; NKN037692V to NKN037698V; NKN037772V to NKN037779V; NKN037790V to NKN037810V; NKN037812V to NKN037840V. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Medtronic Neuromodulation
- Medtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002
FDA Device · Class II · 2026-06-03 · Medtronic Neuromodulation - Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001
FDA Device · Class II · 2026-06-03 · Medtronic Neuromodulation - Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A C…
FDA Device · Class II · 2026-05-20 · Medtronic Neuromodulation - A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200
FDA Device · Class II · 2026-01-21 · Medtronic Neuromodulation - Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US…
FDA Device · Class II · 2025-12-31 · Medtronic Neuromodulation
All Medtronic Neuromodulation recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0567-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.