RecallRadar

FDA Device recall Z-0566-2013

The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of…

On 2012-12-26 the FDA classified a Class II recall of The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of… by Stryker Medical Division, citing model 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the li….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0566-2013
ClassificationClass II
Statusterminated
Initiated2012-10-11
Published2012-12-26
Last updated2026-09-13 11:53 UTC
CompanyStryker Medical Division
DistributionUS

Product

The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of the ambulance.

Reason

Model 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the lifting ability of the arms to be negatively impacted.

Identifiers

code_info
Item 6390000000 Serial Numbers: 120940217 120940257 120940259 120940260 120940261 120940275 120940276 120940277…Item 6390000000 Serial Numbers: 120940217 120940257 120940259 120940260 120940261 120940275 120940276 120940277 120940290 120940291 120940292 120940293 120940294 120940431 120940435 120940436 120940437 120940448 120940449 120940454 120940458 120940459 120940203 120940258

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0566-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.