FDA Device recall Z-0563-2016
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
- FDA Device
- Class II
- completed
- Published 2016-01-06
On 2016-01-06 the FDA classified a Class II recall of Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5 by Trumpf Medical Systems, citing the firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light he….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0563-2016 |
| Classification | Class II |
| Status | completed |
| Initiated | 2015-11-02 |
| Published | 2016-01-06 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Trumpf Medical Systems |
| Distribution | US |
Product
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
Reason
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Identifiers
| code_info | 4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo;…4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo; 4038310 - TruLight trio/quad; 1565068 - iLED single (TUSM); 1565160 - iLED duo (TUSM); 1565161 - iLED trio/quad (TUSM); 1574759 - TruLight single (TUSM); 1574850 - TruLight duo (TUSM); 1574851 - TruLight trio/quad (TUSM) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: completed.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0563-2016%22