RecallRadar

FDA Device recall Z-0562-2015

ASEPT 600 ml Drainage Kit, Ref No. 622287, M7052 Product Usage: The ASEPT Drainage Kit is an accessory to be used to connect to the ASEPT P…

On 2014-12-17 the FDA classified a Class II recall of ASEPT 600 ml Drainage Kit, Ref No. 622287, M7052 Product Usage: The ASEPT Drainage Kit is an accessory to be used to connect to the ASEPT P… by Pfm Medical, citing PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0562-2015
ClassificationClass II
Statusterminated
Initiated2014-10-20
Published2014-12-17
Last updated2026-09-13 11:53 UTC
CompanyPfm Medical
DistributionUS

Product

ASEPT 600 ml Drainage Kit, Ref No. 622287, M7052 Product Usage: The ASEPT Drainage Kit is an accessory to be used to connect to the ASEPT Pleural or Peritoneal Drainage Catheter to drain fluid from the chest or abdomen.

Reason

PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.

Identifiers

code_infoLot No. 1307-065, 1301-009

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0562-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.