FDA Device recall Z-0559-2016
MicroScan Pos Combo Panel Type 33, Part No. B1017-211 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial age…
- FDA Device
- Class II
- terminated
- Published 2016-01-06
On 2016-01-06 the FDA classified a Class II recall of MicroScan Pos Combo Panel Type 33, Part No. B1017-211 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial age… by Beckman Coulter, citing beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organ….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0559-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-12-01 |
| Published | 2016-01-06 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
MicroScan Pos Combo Panel Type 33, Part No. B1017-211 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.
Reason
Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.
Identifiers
| code_info | 2016-04-29 2016-05-04 Part No. B1017-211 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0559-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.