FDA Device recall Z-0554-2024
Sensica Urine Output System, Catalog Number SCCS1002
- FDA Device
- Class II
- ongoing
- Published 2023-12-20
On 2023-12-20 the FDA classified a Class II recall of Sensica Urine Output System, Catalog Number SCCS1002 by C R Bard, citing BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monito….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0554-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-11-15 |
| Published | 2023-12-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | C R Bard |
| Distribution | US |
Product
Sensica Urine Output System, Catalog Number SCCS1002
Reason
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.
Identifiers
| code_info | UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007,…UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007, BMGPRMYS020, BMGPRMYS026, BMGPRMYS028, BMGPRMYS030, BMFWRMSBT4, BMFWRMSJ45, BMFWRMSJ49, BMFZRMS104, BMGNRMS085, BMFSRMSH98, BMFSRMSJ01, BMFVRMSQ62, BMFVRMSQ66, BMFVRMSQ68, BMFYRMSJ41, BMFYRMSJ49, BMFZRMS067, BMFZRMS070, BMFZRMS080, BMFZRMS113, BMFZRMS117, BMGNRMS070, BMFURMSF02, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ89, BMFVRMSQ91, BMFWRSJ46, BMGPRMYS011, BMGPRMYS018, BMGPRMYS022 , BMGPRMYS023, BMGPRMYS027, BMFYRMSJ35, BMFYRMSJ37, BMFYRMSJ54, BMFYRMSJ58, BMGNRMS079, BMGNRMSG69, BMGNRMSG74, BMFSRMSH94, BMFSRMSH99, BMFSRMSJ38, BMFSRMSJ44, BMFTRMS138, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFQRMS248, BMFQRMS250, BMFRRMS646, BMFRRMS647, BMGPRMYS008, BMGPRMYS009, BMGPRMYS010, BMGPRMYS013, BMGPRMYS014, BMGPRMYS016, BMFZRMS073, BMFSRMSH78, BMFSRMSH79, BMFSRMSH80, BMFSRMSH81, BMFSRMSH82, BMFSRMSH83, BM |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by C R Bard
- Products that contain the Pericare Wipes: *Bardex¿ I.C. Complete Care¿ Foley Catheter Tray REF: 300316A UDI codes: 00801741024344¿ 10801741…
FDA Device · Class II · 2026-09-16 · C R Bard - Products that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067¿ 108017411…
FDA Device · Class II · 2026-09-16 · C R Bard - Products that contain the Pericare Wipes: Bard¿ Foley Tray With Lubri-Sil¿ All-Silicone Foley Catheter REF: 947314 947316 947318 UDI codes:…
FDA Device · Class II · 2026-09-16 · C R Bard - Products that contain the Pericare Wipes: (Canada Only) SureStep" Foley Tray System Bardex¿ IC Complete Care¿ REF: A300314AC UDI-DI code: 0…
FDA Device · Class II · 2026-09-16 · C R Bard - Products that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF: INTP16 INTS14 INTS14C INTS16 INTS16C PEDINTS06 PEDINTS08…
FDA Device · Class II · 2026-09-16 · C R Bard
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0554-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.