FDA Device recall Z-0553-2016
GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for…
- FDA Device
- Class II
- terminated
- Published 2018-10-31
On 2018-10-31 the FDA classified a Class II recall of GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for… by Ge Medical Systems, citing coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0553-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-12-07 |
| Published | 2018-10-31 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ge Medical Systems |
| Distribution | AZ, CA, CO, DC, DE, FL, IL, IN, MD, ME, MI, MN, MO, NC, NH, NY, OH, OK, OR, PA, PR, SD, TN, TX, VA, WI |
Product
GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for use on the order of the physician in conjunction with a 3.0T MRI scanner, as an accessory to produce 2D and 3D images.
Reason
Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.
Identifiers
| code_info | Mfg Lot or Serial # System ID 00000001140049 BMDACCMR1 00000001140050 BMDACCMR1 0000011-4-0050 BMDACCMR1 0000011-4-…Mfg Lot or Serial # System ID 00000001140049 BMDACCMR1 00000001140050 BMDACCMR1 0000011-4-0050 BMDACCMR1 0000011-4-0009 920568MR3T 0000011-4-0015 920568MR3T 0000011-4-0017 602406IMR 0000011-4-0018 602406IMR 0000011-4-0024 602406IMR 0000014-4-0007 602406IMR 0000011-4-0028 626359MR3 0000011-4-0032 626359MR3 0000011-4-0046 MSK68MR4 0000011-4-0047 MSK68MR4 0000028-3-0011 MSK68MR4 0000044-4-0046 MSK68MR4 0000011-4-0064 415CB3TW 0000011-4-0065 415CB3TW 0000011-4-0066 415723SHAW12 0000011-4-0067 415723SHAW12 0000011-4-0075 616267IMRI 0000011-4-0076 616267IMRI 011-4-0085 616267IMRI 011-4-0170 616267IMRI 0000011-4-0094 330344MR750W 011-4-0118 330344MR750W 0000011-4-0101 713792CMR1 0000011-4-0102 713792CMR1 0000011-4-0107 317887MR3 0000011-4-0108 317887MR3 0000011-4-0114 309655MR750W 0000011-4-0115 309655MR750W 011-4-0117 517353DVMR 011-4-0172 517353DVMR 011-4-0119 614293MR9 011-4-0120 614293MR9 011-4-0124 MR229050 077-4-0123 MR229050 011-4-0125 6053223TMR |
|---|