Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument…

On 2020-12-16 the FDA classified a Class II recall of Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument… by K2m, citing all-in-One Tab Removal Tool may cause cracking and/or separating at the threaded interface between tool s metal shaft component and handle, resulting in leakage of Loctite from ha….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0542-2021
ClassificationClass II
Statusterminated
Initiated2020-11-04
Published2020-12-16
Last updated2026-09-13 11:55 UTC
CompanyK2m
DistributionUS

Product

Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument back and forth until the entire tab is fully detached from the screw head. Catalog number: 5101-90195

Reason

All-in-One Tab Removal Tool may cause cracking and/or separating at the threaded interface between tool s metal shaft component and handle, resulting in leakage of Loctite from handle cracks

Identifiers

code_infoLot Numbers: JUFT, JUGX, KEFE, KUBJ, KVNT, KYRR, MAJR GTIN: 10888857290679