FDA Device recall Z-0541-2013
Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE E, LEFT TIBIA SIZE E RIGHT This…
- FDA Device
- Class II
- terminated
- Published 2012-12-26
On 2012-12-26 the FDA classified a Class II recall of Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE E, LEFT TIBIA SIZE E RIGHT This… by Zimmer, citing all lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate ar….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0541-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-11-14 |
| Published | 2012-12-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE E, LEFT TIBIA SIZE E RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.
Reason
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
Identifiers
| code_info | Part Number 42-5320-071-01 TIBIA SIZE E, LEFT, Lot 11005332, 11005333, 11005334, 11005359, 11005360, 11005361, 110061…Part Number 42-5320-071-01 TIBIA SIZE E, LEFT, Lot 11005332, 11005333, 11005334, 11005359, 11005360, 11005361, 11006133, 11006134, 11006135, 11006136, 11006137, 11006138, 11006712, 11007124, 61956760, 61956761, 61956762, 61978561, 61978562, 61978563, 62022179, 62030574, 62033389, 62079745, 62030575, 11007596, 62091364, 62103722, 62060744, 62068418, 62079746, 62103723, 62110198, 62117658, 62076580, 62128899, 11007595, 11007597, 62144992, 62150529, 62150530 , Part Number 42-5320-071-02 TIBIA SIZE E RIGHT, Lot 11006515, 11007238, 11007574, 11007575, 11007576, 62018345, 62018346, 62022180, 62022181, 62033391, 62051023, 62051801, 62068419, 62076582, 62079747, 62079761, 62103725, 62110199, 62117659, 62128904, 11007641, 11007640, 11007642, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0541-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.