FDA Device recall Z-0538-2024
AirTouch F20 Full Face Mask and User Guide
- FDA Device
- Class I
- ongoing
- Published 2023-12-27
On 2023-12-27 the FDA classified a Class I recall of AirTouch F20 Full Face Mask and User Guide by Resmed, citing masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0538-2024 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-11-20 |
| Published | 2023-12-27 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Resmed |
| Distribution | US |
Product
AirTouch F20 Full Face Mask and User Guide
Reason
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
Identifiers
| code_info | All mask lots used with User Guide 638220/version 2020-02, and prior. UDI-DI/GTIN: 619498630028, 619498630189, 6194986…All mask lots used with User Guide 638220/version 2020-02, and prior. UDI-DI/GTIN: 619498630028, 619498630189, 619498630172, 619498630141, 619498630059, 619498630097, 619498630080, 619498630011, 619498630004, 619498630042, 619498630066, 619498630134, 619498630073, 619498630035, 619498630158, 619498630165, 619498630523, 619498630530, 619498630516, 619498630363, 619498630356, 619498630349, 619498630387, 619498630370 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0538-2024%22