FDA Device recall Z-0536-2016
ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W
- FDA Device
- Class III
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class III recall of ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0536-2016 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-11-16 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ameditech |
| Distribution | US |
Product
ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W
Reason
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Identifiers
| code_info | 141779 142656 142963 143874 144562 151996 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0536-2016%22