RecallRadar

FDA Device recall Z-0536-2014

MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of…

On 2013-12-25 the FDA classified a Class II recall of MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of… by Roche Diagnostics Operations, citing this issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch wh….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0536-2014
ClassificationClass II
Statusterminated
Initiated2013-11-25
Published2013-12-25
Last updated2026-09-13 11:53 UTC
CompanyRoche Diagnostics Operations
DistributionAL, AZ, CA, CT, GA, NC, NJ, NY, OH, TX, VA, WA

Product

MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) by the use of the specially designed MagNA Pure reagent kits, automated filling of different kinds of PCR reaction vessels.

Reason

This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD instrument (catalog number 06541089001). Importing an order file in xml format to the MagNA Pure 96 IVD instrument software may result in a wrong sequence of samples in the graphical overvi

Identifiers

code_info
Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 20…Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 2024, 2025, 2026, 2027, 2028, 4001, 4022, 4023, 4024, 4025,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0536-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.