RecallRadar

FDA Device recall Z-0532-2018

Spacelabs Healthcare Xhibit Telemetry Receiver, Model 96280, software version 1.1, physiological patient monitor (with arrhythmia detection…

On 2018-02-14 the FDA classified a Class II recall of Spacelabs Healthcare Xhibit Telemetry Receiver, Model 96280, software version 1.1, physiological patient monitor (with arrhythmia detection… by Spacelabs Healthcare, citing spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0532-2018
ClassificationClass II
Statusterminated
Initiated2018-01-17
Published2018-02-14
Last updated2026-09-13 11:54 UTC
CompanySpacelabs Healthcare
DistributionGA, MI, NJ, VA

Product

Spacelabs Healthcare Xhibit Telemetry Receiver, Model 96280, software version 1.1, physiological patient monitor (with arrhythmia detection or alarms) The Spacelabs Healthcare Telemetry Receiver, Model 96280, is intended to provide the Spacelabs Healthcare monitoring system with adult, pediatric and neonatal patient data of patients connected to Spacelabs Healthcare telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demographic information to monitor adequacy of treatment or to exclude causes of symptoms.

Reason

Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.

Identifiers

code_info
Serial numbers: 6280-000359, 6280-000340, 6280-000336, 6280-000503, 6280-000502, 6280-000500, 6280-000496, 6280-000495…Serial numbers: 6280-000359, 6280-000340, 6280-000336, 6280-000503, 6280-000502, 6280-000500, 6280-000496, 6280-000495, 6280-000327, 6280-000260, 6280-000566, 6280-000290

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0532-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.