FDA Device recall Z-0529-2019
Gel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat
- FDA Device
- Class III
- terminated
- Published 2018-12-12
On 2018-12-12 the FDA classified a Class III recall of Gel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat by Helena Laboratories, citing due to a customer experiencing complaints on bone fraction of assay not being in range. The complaint: ~12% lower enzyme activity was noted.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0529-2019 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2016-11-07 |
| Published | 2018-12-12 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Helena Laboratories |
| Distribution | US |
Product
Gel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat. No. 5104 1 x 2.0 mL FOR IN-VITRO DIAGNOSTIC USE Gel Alkaline Phosphatase Isoenzyme Control 20mL 001831 Alkaline Phosphatase Isoenzyme Control is to be used as a qualitative and/or quantitative control to aid in the identification of alkaline Phosphatase isoenzymes by electrophoresis.
Reason
Due to a customer experiencing complaints on bone fraction of assay not being in range. The complaint: ~12% lower enzyme activity was noted.
Identifiers
| code_info | " Lot Codes: 1-16-5104 (Scheme: 1(st lot)-(of 20)16-(for C/N )5104) " Model/Catalog Number: 5104 " UDI: Primary DI - M52551040 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Other recalls by Helena Laboratories
- Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30 Blade Applicators. used in invitr…
FDA Device · Class II · 2023-06-28 · Helena Laboratories - V8 Immunodisplacement Kit REF 1803
FDA Device · Class II · 2022-06-15 · Helena Laboratories - Cascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo analyzer and are intended for the determination of th…
FDA Device · Class II · 2021-02-17 · Helena Laboratories - K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING TIME ANALYZER, and ACTALYKE MINI ACTIVATED CLOTTING…
FDA Device · Class II · 2020-12-23 · Helena Laboratories - Collagen Reagent, Cat
FDA Device · Class III · 2019-12-04 · Helena Laboratories