FDA Device recall Z-0526-2022
1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C de…
- FDA Device
- Class II
- ongoing
- Published 2022-02-02
On 2022-02-02 the FDA classified a Class II recall of 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C de… by Nrt X Ray A S, citing friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to us….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0526-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-01-21 |
| Published | 2022-02-02 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Nrt X Ray A S |
| Distribution | IA, KY, LA, MA, NC, NY, OH, SC, WA, WV |
Product
1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)
Reason
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
Identifiers
| code_info | 1. Footrest, REF: 03282000, UDI: 05713464000336; Serial Numbers: 903282000046, 903282000057, 903282000075 ,903282000…1. Footrest, REF: 03282000, UDI: 05713464000336; Serial Numbers: 903282000046, 903282000057, 903282000075 ,9032820000070, 903282000001, 903282000002, 903282000004 thru 903282000006, 903282000009, 903282000011, 903282000012 903282000015, 903282000016, 903282000018 thru 903282000020, 903282000023, 903282000024, 903282000025, 903282000026, 903282000028, 903282000032 thru 903282000034, 903282000036, 903282000038 thru 903282000042, 903282000044, 903282000045, 903282000047 thru 903282000051, 903282000053, 903282000055, 903282000056, 903282000058 thru 903282000061, 903282000063, 903282000064, 903282000067, 903282000068 2. Shoulder Support, REF: 03283000, UDI: 05713464000381 Serial Numbers: 903283000012, 903283000028, 903283000001 thru 903283000003, 903283000005, 903283000010, 903283000011, 903283000013 thru 903283000022, 903283000023, 903283000024, 903283000026 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0526-2022%22