RecallRadar

FDA Device recall Z-0525-2013

Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-cont…

On 2013-01-02 the FDA classified a Class I recall of Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-cont… by Praxair, citing isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0525-2013
ClassificationClass I
Statusterminated
Initiated2012-11-16
Published2013-01-02
Last updated2026-09-13 11:53 UTC
CompanyPraxair
DistributionUS

Product

Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.

Reason

Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Identifiers

code_info
OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal…OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed.