RecallRadar

FDA Device recall Z-0518-2020

Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body

On 2019-12-04 the FDA classified a Class II recall of Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body by Becton Dickinson And, citing the indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0518-2020
ClassificationClass II
Statusterminated
Initiated2019-11-04
Published2019-12-04
Last updated2026-09-13 11:55 UTC
CompanyBecton Dickinson And
DistributionUS

Product

Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite¿ 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite¿ 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Console with Connector, Catalog No.: 9770600 Site~Rite 8 Ultrasound System Console with Connector, Refurbished, Catalog No: 9770600X Site~Rite 8 Ultrasound System Console with Connector (Remanufactured) Catalog No: 9770600Y

Reason

The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only

Identifiers

code_infoAll Serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0518-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.