- Published 2019-12-04
On 2019-12-04 the FDA classified a Class II recall of Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body by Becton Dickinson And, citing the indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0518-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-11-04 |
| Published | 2019-12-04 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite¿ 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite¿ 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Console with Connector, Catalog No.: 9770600 Site~Rite 8 Ultrasound System Console with Connector, Refurbished, Catalog No: 9770600X Site~Rite 8 Ultrasound System Console with Connector (Remanufactured) Catalog No: 9770600Y
Reason
The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only
Identifiers
| code_info | All Serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0518-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.