Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC…
- FDA Device
- Class II
- terminated
- Published 2020-12-09
On 2020-12-09 the FDA classified a Class II recall of Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC… by Invacare, citing breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0514-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-10-30 |
| Published | 2020-12-09 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Invacare |
| Distribution | US |
Product
Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5
Reason
Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.
Identifiers
| code_info | Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633;…Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633; IRC5PO2VAW from serial #17HF030338-18IF018522; IRC5PO2VAWL2 from serial # 18BF007341-18IF024451; IRC5PO2VAWL3 from serial # 17JF002563-19AF008559; IRC5PO2VAWL5 from serial # 18EF003466-18EF003544. DI Number: 00841447100249 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Other recalls by Invacare
- Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3) G54 (4) BAR5490IVC (5) BAR5…
FDA Device · Class II · 2024-03-13 · Invacare - DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat
FDA Device · Class II · 2023-08-30 · Invacare - Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)
FDA Device · Class III · 2022-07-20 · Invacare - Invacare TDX SP2, Model Nos
FDA Device · Class II · 2022-06-01 · Invacare - Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R…