RecallRadar

FDA Device recall Z-0511-2016

ImmuTest 12 Panel Dip Card, Item No. IMD-12MO3

On 2015-12-30 the FDA classified a Class III recall of ImmuTest 12 Panel Dip Card, Item No. IMD-12MO3 by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0511-2016
ClassificationClass III
Statusterminated
Initiated2015-11-16
Published2015-12-30
Last updated2026-09-13 11:53 UTC
CompanyAmeditech
DistributionUS

Product

ImmuTest 12 Panel Dip Card, Item No. IMD-12MO3

Reason

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Identifiers

code_info144009 144078 144099 144125 144193 144268 152996 153077

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.