FDA Device recall Z-0511-2013
Siemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Int…
- FDA Device
- Class II
- terminated
- Published 2012-12-19
On 2012-12-19 the FDA classified a Class II recall of Siemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Int… by Siemens Healthcare Diagnostics, citing operating range of the ADVIA Centaur systems is 18C¿ to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific ass….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0511-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-11-14 |
| Published | 2012-12-19 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
Siemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Intended Use For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems.
Reason
Operating range of the ADVIA Centaur systems is 18C¿ to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Identifiers
| code_info | Al lot codes |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0511-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.