FDA Device recall Z-0509-2013

Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for…

On 2012-12-19 the FDA classified a Class II recall of Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for… by Siemens Healthcare Diagnostics, citing operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0509-2013
ClassificationClass II
Statusterminated
Initiated2012-11-14
Published2012-12-19
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for the quantitative determination of dehydroepiandrosterone sulfate (DHEAS) in human serum and plasma using the ADVIA Centaur and ADVIA Centaur XP systems

Reason

Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Identifiers

code_infoAll lot codes

Status history