Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5

On 2015-12-30 the FDA classified a Class III recall of Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5 by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0507-2016
ClassificationClass III
Statusterminated
Initiated2015-11-16
Published2015-12-30
Last updated2026-09-13 11:53 UTC
CompanyAmeditech
DistributionUS

Product

Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5

Reason

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.