FDA Device recall Z-0505-2026
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US
- FDA Device
- Class II
- ongoing
- Published 2025-11-26
On 2025-11-26 the FDA classified a Class II recall of Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US by Thoratec, citing due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0505-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-10-09 |
| Published | 2025-11-26 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Thoratec |
| Distribution | US |
Product
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No
Reason
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Identifiers
| code_info | Lot Code: All units of the following models are impacted: Model 106524US UDI: 00813024013297 Model 106524INT UDI: 0081…Lot Code: All units of the following models are impacted: Model 106524US UDI: 00813024013297 Model 106524INT UDI: 0081302401171 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used. |
|---|