FDA Device recall Z-0501-2016
EDI BZO Uncut Sheet - Cup*, Item No. 160170
- FDA Device
- Class III
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class III recall of EDI BZO Uncut Sheet - Cup*, Item No. 160170 by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0501-2016 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-11-16 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ameditech |
| Distribution | US |
Product
EDI BZO Uncut Sheet - Cup*, Item No. 160170
Reason
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Identifiers
| code_info | 141260 141294 141909 141925 142074 142243 142428 142614 142712 142911 144555 151925 152003 152223 152498…141260 141294 141909 141925 142074 142243 142428 142614 142712 142911 144555 151925 152003 152223 152498 152723 152897 153311 153559 153818 154184 152103 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0501-2016%22