FDA Device recall Z-0499-2022
1. Sensis, material # 10764561
- FDA Device
- Class II
- ongoing
- Published 2022-01-26
On 2022-01-26 the FDA classified a Class II recall of 1. Sensis, material # 10764561 by Siemens Medical Solutions Usa, citing software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0499-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-11-23 |
| Published | 2022-01-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.
Reason
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.
Identifiers
| code_info | 1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11…1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 1030 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0499-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.