FDA Device recall Z-0492-2018
Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932
- FDA Device
- Class II
- terminated
- Published 2018-02-07
On 2018-02-07 the FDA classified a Class II recall of Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932 by Tosoh Bioscience, citing this recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0492-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-06-11 |
| Published | 2018-02-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Tosoh Bioscience |
| Distribution | US |
Product
Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Reason
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Identifiers
| code_info | AIA-900 19tray Sorter with S/N 10000109 to S/N 80136909. AIA-900 9tray Sorter with S/N 10000109 to S/N 10707402. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0492-2018%22