RecallRadar

FDA Device recall Z-0492-2018

Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932

On 2018-02-07 the FDA classified a Class II recall of Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932 by Tosoh Bioscience, citing this recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0492-2018
ClassificationClass II
Statusterminated
Initiated2017-06-11
Published2018-02-07
Last updated2026-09-13 11:54 UTC
CompanyTosoh Bioscience
DistributionUS

Product

Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.

Reason

This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.

Identifiers

code_infoAIA-900 19tray Sorter with S/N 10000109 to S/N 80136909. AIA-900 9tray Sorter with S/N 10000109 to S/N 10707402.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.