FDA Device recall Z-0492-2016
DrugCheck Waived 12 Panel Cup w/Adult, Item No. DCC-81205-5
- FDA Device
- Class III
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class III recall of DrugCheck Waived 12 Panel Cup w/Adult, Item No. DCC-81205-5 by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0492-2016 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-11-16 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ameditech |
| Distribution | US |
Product
DrugCheck Waived 12 Panel Cup w/Adult, Item No. DCC-81205-5
Reason
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Identifiers
| code_info | 143235 144216 151591 153132 153489 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0492-2016%22