FDA Device recall Z-0489-2018
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the terminatio…
- FDA Device
- Class I
- terminated
- Published 2018-02-14
On 2018-02-14 the FDA classified a Class I recall of MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the terminatio… by Philips North America, citing affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0489-2018 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2017-07-12 |
| Published | 2018-02-14 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Philips North America |
| Distribution | n/a |
Product
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Reason
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Identifiers
| code_info | Serial Numbers: US00601802, US00601968, US00601969, US00602114, US00602122, US00602124, US00602125, US00602126, US006…Serial Numbers: US00601802, US00601968, US00601969, US00602114, US00602122, US00602124, US00602125, US00602126, US00602127, US00602128, US00602145, US00602180, US00602182, US00602184, US00602185, US00602186, US00602187, US00602189, US00602206, US00602236, US00602237, US00602238, US00602239, US00602240, US00602241, US00602242, US00602243, US00602244, US00602245, US00602246, US00602247, US00602248, US00602250, US00602251, US00602252, US00602253, US00602254, US00602344, US00602345, US00602346, US00602347, US00602348, US00602349, US00602350, US00602351, US00602352, US00602353, US00602354, US00602355, US00602356, US00602357, US00602358, US00602359, US00602360, US00602361, US00602362, US00602363, US00602364, US00602365, US00602366, US00602367, US00602368, US00602369, US00602370, US00602371, US00602372, US00602373, US00602374, US00602375, US00602376, US00602377 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0489-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.