RecallRadar

FDA Device recall Z-0488-2016

12-Panel Dip Drug Screen with adulterant, Item No. PSDA-12BUP, PSD-6MTDBO300, PSD-10PPX, PSD-10MOX, 60960D, 61127D, DCC-81205-5, IMCA-10M5

On 2015-12-30 the FDA classified a Class III recall of 12-Panel Dip Drug Screen with adulterant, Item No. PSDA-12BUP, PSD-6MTDBO300, PSD-10PPX, PSD-10MOX, 60960D, 61127D, DCC-81205-5, IMCA-10M5 by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0488-2016
ClassificationClass III
Statusterminated
Initiated2015-11-16
Published2015-12-30
Last updated2026-09-13 11:53 UTC
CompanyAmeditech
DistributionUS

Product

12-Panel Dip Drug Screen with adulterant, Item No. PSDA-12BUP, PSD-6MTDBO300, PSD-10PPX, PSD-10MOX, 60960D, 61127D, DCC-81205-5, IMCA-10M5.

Reason

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Identifiers

code_info151688 152423 153307 153912 153921 153961 153962 154104 154112 154236 154264

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0488-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.