Ascom Telligence Nurse Call System

On 2018-11-28 the FDA classified a Class II recall of Ascom Telligence Nurse Call System by Ascom Us, citing there is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights con….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0486-2019
ClassificationClass II
Statusongoing
Initiated2018-11-02
Published2018-11-28
Last updated2026-09-13 11:54 UTC
CompanyAscom Us
DistributionAL, CA, CO, GA, IN, KY, LA, ME, MI, MN, MO, MS, NE, NV, OH, PA, RI, TX, UT, VA, WI

Product

Ascom Telligence Nurse Call System

Reason

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Identifiers

code_infoVersion 5.0 (Build 5.0.4)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.