FDA Device recall Z-0485-2016
10-Panel Pipette Drug Screen, Item No. PSP-10M
- FDA Device
- Class III
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class III recall of 10-Panel Pipette Drug Screen, Item No. PSP-10M by Ameditech, citing ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0485-2016 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2015-11-16 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ameditech |
| Distribution | US |
Product
10-Panel Pipette Drug Screen, Item No. PSP-10M
Reason
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Identifiers
| code_info | 153135 154119 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0485-2016%22