RecallRadar

FDA Device recall Z-0480-2026

GM85 Digital Mobile X-ray imaging System

On 2025-11-19 the FDA classified a Class II recall of GM85 Digital Mobile X-ray imaging System by Neurologica, citing the four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0480-2026
ClassificationClass II
Statusongoing
Initiated2025-10-07
Published2025-11-19
Last updated2026-09-13 11:57 UTC
CompanyNeurologica
DistributionUS

Product

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Reason

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Identifiers

code_info
Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IH…Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0480-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.