FDA Device recall Z-0480-2014
NexGen¿ Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constraine…
- FDA Device
- Class II
- terminated
- Published 2013-12-18
On 2013-12-18 the FDA classified a Class II recall of NexGen¿ Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constraine… by Zimmer, citing two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial p….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0480-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-31 |
| Published | 2013-12-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
NexGen¿ Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis. The device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis; Collagen disorders, and/or avascular necrosis of the femoral condyle; Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, varus, or flexion deformities.
Reason
Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial plate would not hold the stem as the connection was too loose. The taper diameter was found to be oversized by .010.
Identifiers
| code_info | 00-5980-037-02; lot 6228417 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Zimmer
- Off-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040476
FDA Device · Class II · 2026-07-01 · Zimmer - Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241
FDA Device · Class II · 2026-07-01 · Zimmer - Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240
FDA Device · Class II · 2026-07-01 · Zimmer - Brand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/Catalog Number: 42-5572-066-10…
FDA Device · Class II · 2026-07-01 · Zimmer - Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
FDA Device · Class II · 2026-07-01 · Zimmer
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0480-2014%22