RecallRadar

FDA Device recall Z-0478-2019

Insulia Diabetes Management Companion

On 2018-11-28 the FDA classified a Class II recall of Insulia Diabetes Management Companion by Voluntis, citing the firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0478-2019
ClassificationClass II
Statusterminated
Initiated2018-10-08
Published2018-11-28
Last updated2026-09-13 11:54 UTC
CompanyVoluntis
DistributionAZ, MI, OH, TN, VT

Product

Insulia Diabetes Management Companion

Reason

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

Identifiers

code_infoSoftware version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0478-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.