FDA Device recall Z-0474-2023
Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187
- FDA Device
- Class II
- terminated
- Published 2022-12-21
On 2022-12-21 the FDA classified a Class II recall of Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187 by Conmed, citing suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could ca….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0474-2023 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-11-18 |
| Published | 2022-12-21 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Conmed |
| Distribution | US |
Product
Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187
Reason
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Identifiers
| code_info | UDI-DI: 30653405059930 Lot codes 202204294 202205134 202205164 202206084 202206154 202206204 202206304 202207074 202207084 202208014 202208044 202208084 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0474-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.