RecallRadar

FDA Device recall Z-0473-2013

Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001

On 2012-12-12 the FDA classified a Class II recall of Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001 by Roche Diagnostics Operations, citing on cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditio….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0473-2013
ClassificationClass II
Statusterminated
Initiated2012-11-01
Published2012-12-12
Last updated2026-09-13 11:53 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.

Reason

On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C

Identifiers

code_info
Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC wi…Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff¿ Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05