FDA Device recall Z-0472-2023
ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear…
- FDA Device
- Class II
- ongoing
- Published 2022-12-21
On 2022-12-21 the FDA classified a Class II recall of ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear… by Siemens Medical Solutions Usa, citing when selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0472-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-11-09 |
| Published | 2022-12-21 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
Reason
When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter
Identifiers
| code_info | UDI: N/A S/N: 10736, 10737, 10822, 10850, 10886,10843, 10865,10025 Software version 4.3.1 MR3 |
|---|