FDA Device recall Z-0471-2022
Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue
- FDA Device
- Class II
- terminated
- Published 2022-01-19
On 2022-01-19 the FDA classified a Class II recall of Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue by Philips North America, citing (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0471-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-11-12 |
| Published | 2022-01-19 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency stop recovery issue Model Numbers: 722221, 722228, 722225, 722068, 722226, 722078, 722224
Reason
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
Identifiers
| code_info | US: Model 722068 Azurion 7 B20 S/N: 109 UDI: (01)00884838085367(21)109. OUS: Model:…US: Model 722068 Azurion 7 B20 S/N: 109 UDI: (01)00884838085367(21)109. OUS: Model: S/N UDI: 722226 Azurion 7 B20 176 (01)00884838099272(21)176; 722228 Azurion 5 M20 3 (01)00884838099234(21)3; 722221 Azurion 3 M12 8 (01)00884838099203(21)8; 722225 Azurion 7 B12 16 (01)00884838099265(21)16; 722225 Azurion 7 B12 73 (01)00884838099265(21)73; 722226 Azurion 7 B20 42 (01)00884838099272(21)42; 722226 Azurion 7 B20 157 (01)00884838099272(21)157; 722224 Azurion 7 M20 293 (01)00884838099258(21)293; 722078 Azurion 7 M12 220 (01)00884838085251(21)220 |
|---|