FDA Device recall Z-0468-2019
Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Produc…
- FDA Device
- Class II
- terminated
- Published 2018-11-21
On 2018-11-21 the FDA classified a Class II recall of Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Produc… by Arrow International, citing the products labeling contains inconsistencies. In certain instances, the lidstock, banner card and shipper label state that the Spring Wire Guide (SWG) within is .032 rather than….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0468-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-09-21 |
| Published | 2018-11-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Arrow International |
| Distribution | US |
Product
Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Product Usage: The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg+ard Blue Plus antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use.
Reason
The products labeling contains inconsistencies. In certain instances, the lidstock, banner card and shipper label state that the Spring Wire Guide (SWG) within is .032 rather than the correct SWG size .025 . For other lots in scope the lidstock correctly states the SWG size .025 however the banner card states the incorrect SWG size .032 .
Identifiers
| code_info | Lot/Batch Numbers: 13F17E0685 13F17F0562 13F17H0141 13F17K0340 13F18B0119 13R17E0685 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0468-2019%22