FDA Device recall Z-0463-2013

Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1…

On 2012-12-12 the FDA classified a Class II recall of Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1… by Sekisui Diagnostics, citing spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0463-2013
ClassificationClass II
Statusterminated
Initiated2012-10-24
Published2012-12-12
Last updated2026-09-13 11:53 UTC
CompanySekisui Diagnostics
DistributionUS

Product

Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1 in human plasma. Catalog # 101201. Spectrolyse PAI-1 is intended for the quantitative determination of Plasminogen Activator Inhibitor Type-1 (PAI-1) activity in human plasma. The test is for in vitro diagnostic use.

Reason

Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.

Identifiers

code_infoLot # 120827, exp 11/27/2012