FDA Device recall Z-0454-2013
Neptune¿ 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max…
- FDA Device
- Class II
- terminated
- Published 2012-12-12
On 2012-12-12 the FDA classified a Class II recall of Neptune¿ 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max… by Stryker Instruments Div, citing beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0454-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-05-24 |
| Published | 2012-12-12 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Instruments Div |
| Distribution | US |
Product
Neptune¿ 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.
Reason
Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.
Identifiers
| code_info | Part 0702-001-000 (120v) Lot 0733700353 through 1036100423 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0454-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.