FDA Device recall Z-0452-2022
Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555
- FDA Device
- Class II
- terminated
- Published 2022-01-12
On 2022-01-12 the FDA classified a Class II recall of Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555 by Biomeme, citing the current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to short….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0452-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-10-20 |
| Published | 2022-01-12 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Biomeme |
| Distribution | AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA |
Product
Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564
Reason
The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).
Identifiers
| code_info | All lots. Last lot of Kit distributed with the old IFU is 20210914. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0452-2022%22