FDA Device recall Z-0450-2026
Convenience kits
- FDA Device
- Class I
- ongoing
- Published 2025-11-19
On 2025-11-19 the FDA classified a Class I recall of Convenience kits by Medline Industries, citing firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and ina….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0450-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-10-10 |
| Published | 2025-11-19 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
Reason
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Identifiers
| code_info | GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYN…GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0450-2026%22