FDA Device recall Z-0450-2016
A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconst…
- FDA Device
- Class II
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class II recall of A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconst… by Biomet, citing these screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0450-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-20 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Biomet |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
Product
A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.
Reason
These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.
Identifiers
| code_info | Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16…Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16MM; 816140018, 4.0X18MM; 816140020, 4.0X20MM; 816140022, 4.0X22MM; 816140024, 4.0X24MM; 816140026, 4.0X26MM; 816140028, 4.0X28MM; 816140030, 4.0X30MM; 816140032, 4.0X32MM; 816140034, 4.0X34MM; 816140036, 4.0X36MM; 816140038, 4.0X38MM; 816140040, 4.0X40MM; 816140042, 4.0X42MM; 816140044, 4.0X44MM; 816140046, 4.0X46MM; 816140048, 4.0X48MM; 816140050, 4.0X50MM; 816140055, 4.0X55MM; 816140060, 4.0X60MM; 816140065, 4.0X65MM; 816140070, 4.0X70MM; 816140075, 4.0X75MM; 816140080, 4.0X80MM; 816140085, 4.0X85MM; 816140090, 4.0X90MM; 816140095, 4.0X95MM; 816140100, 4.0X100MM Part number and size (sterile) 856140010, 4.0X10MM; 856140012, 4.0X12MM; 856140014, 4.0X14MM; 856140016, 4.0X16MM; 856140018, 4.0X18MM; 856140020, 4.0X20MM; 856140022, 4.0X22MM; 856140024, 4.0X24MM; 856140026, 4.0X26MM; 856140028, 4.0X28MM; 856140030, 4.0X30MM; 856140 |
|---|