RecallRadar

FDA Device recall Z-0447-2021

Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of…

On 2020-11-18 the FDA classified a Class II recall of Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of… by Shanghai United Imaging Healthcare, citing A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0447-2021
ClassificationClass II
Statusongoing
Initiated2020-08-05
Published2020-11-18
Last updated2026-09-13 11:55 UTC
CompanyShanghai United Imaging Healthcare
DistributionUS

Product

Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Identifiers

code_infoSerial Numbers: 600229, 600230, 600232, 600271, 600282

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0447-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.