ELEKTA LIMITED Versa HD, REF XRT 2121, 2131

On 2023-12-13 the FDA classified a Class II recall of ELEKTA LIMITED Versa HD, REF XRT 2121, 2131 by Elekta, citing elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0445-2024
ClassificationClass II
Statusongoing
Initiated2023-10-25
Published2023-12-13
Last updated2026-09-13 11:56 UTC
CompanyElekta
Distributionn/a

Product

ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Reason

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Identifiers

code_info
UDI/DI 05060191071529, Serial Numbers: 153396, 153671, 109003, 153383, 153384, 153055, 153980, 154001, 154002, 154003,…UDI/DI 05060191071529, Serial Numbers: 153396, 153671, 109003, 153383, 153384, 153055, 153980, 154001, 154002, 154003, 154004, 154272, 153104, 153188, 153243, 153745, 153836, 153872, 153876, 154123, 154148, 154199, 154250, 154260, 154267, 154269, 154449, 154498, 154550, 152951, 153140, 153525, 153526, 153532, 154218, 153621, 153951, 153949, 154028, 154548, 154239, 154262, 153413, 153829, 154163, 154101, 152777, 153037, 153142, 153144, 153177, 153186, 153187, 153189, 153277, 153364, 153399, 153425, 153426, 153471, 153510, 153537, 153578, 153625, 153661, 153752, 153762, 153782, 153797, 153808, 153816, 153847, 153858, 153950, 153986, 153997, 154046, 154070, 154084, 154085, 154127, 154178, 154202, 154217, 154280, 154372, 154374, 154381; UDI/DI 05060191071574, Serial Numbers: 153722, 109005, 109007, 153003, 153008, 153314, 153337, 153427, 153428, 153482, 153649, 153900, 154023, 154024, 154050, 154087, 154299, 154300, 153248, 153699, 154043, 154395, 154435, 154179, 154131, 153658, 153830