FDA Device recall Z-0445-2022
Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP…
- FDA Device
- Class II
- ongoing
- Published 2022-01-12
On 2022-01-12 the FDA classified a Class II recall of Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP… by Invacare, citing power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0445-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-11-12 |
| Published | 2022-01-12 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Invacare |
| Distribution | US |
Product
Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG
Reason
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Identifiers
| code_info | LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: TDXSP2 00841447100911 TDXSP2-CG 00841447100942 TDXSP2…LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: TDXSP2 00841447100911 TDXSP2-CG 00841447100942 TDXSP2-MCG 00841447100966 TDXSP2HD 00841447108641 TDXSP2HD-CG 00841447111207 TDXSP2HD-MCG 00841447111214 TDXSP2V 00841447101512 TDXSP2V-HD 00841447102052 TDXSP2X-CG 00841447101000 TDXSP2X-MCG 00841447101017 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0445-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.