FDA Device recall Z-0445-2014
GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc
- FDA Device
- Class II
- terminated
- Published 2013-12-11
On 2013-12-11 the FDA classified a Class II recall of GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc by Smith And Nephew, citing the Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0445-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-08-27 |
| Published | 2013-12-11 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Smith And Nephew |
| Distribution | NV |
Product
GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
Reason
The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.
Identifiers
| code_info | Batch No. 12GT20824 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Smith And Nephew
- TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
FDA Device · Class II · 2026-06-24 · Smith And Nephew - BIOSURE HA SCREW, 6MM x 20 MM. Threaded metallic bone fixation fastener
FDA Device · Class II · 2025-12-24 · Smith And Nephew - META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
FDA Device · Class II · 2025-05-28 · Smith And Nephew - BIORAPTOR Suture Anchors
FDA Device · Class II · 2024-12-11 · Smith And Nephew - Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar co…
FDA Device · Class II · 2024-11-13 · Smith And Nephew
Source