FDA Device recall Z-0443-2024
ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511
- FDA Device
- Class II
- ongoing
- Published 2023-12-13
On 2023-12-13 the FDA classified a Class II recall of ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511 by Elekta, citing elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0443-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-25 |
| Published | 2023-12-13 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Elekta |
| Distribution | n/a |
Product
ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Reason
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Identifiers
| code_info | UDI/DI 05060191071505, Serial Numbers: 154549, 151917, 151963, 152072, 152081, 152131, 152382, 151838, 152318, 152807,…UDI/DI 05060191071505, Serial Numbers: 154549, 151917, 151963, 152072, 152081, 152131, 152382, 151838, 152318, 152807, 151951, 152416, 153147, 153148, 153963, 151821, 151845, 151940, 152160, 152162, 152164, 152389, 152393, 152394, 152634, 152635, 152636, 152664, 152665, 152725, 152992, 151850, 151920, 152082, 152089, 152111, 152546, 152548, 152299, 152403, 152503, 152676, 152950, 151857, 151690, 151744, 151745, 151766, 151831, 151859, 151900, 151950, 152046, 152137, 152194, 152202, 152216, 152258, 152276, 152376, 152390, 152391, 152399, 152414, 152430, 152431, 152513, 152521, 152557, 152576, 152645, 152647, 152712, 152719, 152740, 152743, 153093, 153115, 153126, 153258, 153299, 153475, 153554, 153804, 153911, 154415, 151852, 151894, 151971, 152185, 152241, 152297, 152422, 152429, 152584, 152616, 152628, 152631, 152681, 152698, 152746, 152747, 152770, 152804, 152811, 152812, 152813, 152985, 153256, 153286, 153546, 152014, 152208, 153952, 152883, 152714, 154342, 152724, 151366, 152047, 1 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Elekta
- Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator Product Description: The Flet…
FDA Device · Class II · 2026-07-15 · Elekta - Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced…
FDA Device · Class II · 2026-07-15 · Elekta - Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for intracavitary brachytherapy for c…
FDA Device · Class II · 2026-07-15 · Elekta - MOSAIQ Oncology Information System
FDA Device · Class II · 2026-06-17 · Elekta - Brand Name: Leksell Vantage Arc System REF: 1053958
FDA Device · Class II · 2026-05-06 · Elekta
Source